Build Your Online Product Catalogs?



Description
Product Name: Dielectric loss measurement equipment
Supply Ability:
Related proudcts frequency conversion, high accuracy,
Specifications CE ISO9001 AND SO ON
Price Term: FOB,CNF,CIF,Other
Port of loading: SHENZHEN
Minimum Order 1
Unit Price: $5100~$5130

XHJS1000B Dielectric loss measurement equipment

XHJS1000B dielectric measurement equipment is equipped with thermal printer on the basis of XHJS1000A dielectric loss tester; add bad Loop discharge prompts; enhanced anti-masking function of low voltage wiring, can simultaneously measure capacity and dielectric dissipation values of the two capacitors; also can measure CVT ratio and phase error; output voltage up to 12kV.
Technical Indicate
Accuracy: Cx: ±(Reading×1%+1pF)
tgδ: ±(Reading×1%+0.00040)
Anti interference index: frequency conversion anti interference, under the 200% of interference can still achieve the accuracy of the above
Capacitance range:
Internal high voltage : 3pF~60000pF/10kV 60pF~1μF/0.5kV ; External high voltage:3pF~1.5μF/10kV 60pF~30μF/0.5kV
Resolution:Max. 0.001pF, 4 effective numbers
tgδrange:Unlimited,Resolution 0.001%, Automatic identification of three kinds of test items for capacitance, inductance and resistance
test current range:10μA~1A
Internal high voltage:Set voltage range:0.5~10kV
Max output current:200mA
Lifting and pressing mode: continuous and smooth adjustment
Test frequency:45,50,55,60,65Hz single frequency
45/55Hz,55/65Hz,47.5/52.5Hz Automatic double frequency conversion
Frequency accuracy: ±0.01Hz
External high voltage: 40-70Hz Under positive connection, the maximum test current 1A, power frequency or frequency conversion 40-70Hz
Under reverse connection, the maximum test current 10kV/1A, power frequency or frequency conversion 40-70Hz
CVT self excitation method low voltage output: output voltage 3 ~ 50V, output current 3 ~ 30A
CVT turns ratio test:
Turns ratio test accuracy:±reading×1%
Turns ratio test range: 10~99999
Phase test accuracy: ±0.1° Phase test range: 0~359.9°
Test time: about 40s, that related to test method
Input power: 180V~270VAC,50Hz±1%,electricity or Generator power supply
Computer interface: Standard RS232 interface
Printer: Weihuang A7 micro thermal printer
Temperature:-10℃~50�
Relative humidity

Contact Us
Company: Xi'an Xu & Hui Electromechanical Technology Co.,Ltd.
Contact: Ms. Daisy Yang
Address: Room 704, Dongxing Technology Building, Xincheng Technology Industrial Park, Xincheng District, Xi'an,Shaanxi, China
Postcode:
Tel: 86 29 82565867
Fax: 86 29 82365090
E-mail:         


Copyright© Xi'an Xu & Hui Electromechanical Technology Co.,Ltd. All Rights Reserved.
Tel : 86 29 82565867 Fax : 86 29 82365090
Powered by 86exporter.com